Welcome to Bashitu Medica. We provides Exclusive medical accessories

Double Channel Syringe Pump BYZ-810T

Model: BYZ-810T
Manu: Changsha BEYOND Medical Devices Co.,
Ltd.
Origin: China.

Reviews

There are no reviews yet.

Be the first to review “Double Channel Syringe Pump BYZ-810T”

Your email address will not be published. Required fields are marked *

Description

Features of Dual Channel Syringe Pump
1. HD LCD Display, high capacity words, friendly user interface, dynamically display working status;
2. Visual alarm for occlusion, empty, near empty, low battery, end of infusion, syringe loose, wrong setting, etc.;
3. The syringe pump can be operated separately and compatible with 5ml, 10ml, 20ml, 30ml, 50/60ml syringe of any brands;
4. Preset solution volume to greatly reduce the workload of nurses;
5. Three working modes: Rate mode, Time Volume mode, Dosage Weight mode;
6. Three levels of occlusion: high, medium and low;
7. Purge and Bolus function;
8. KVO (keep-vein-open) automatically opens as injection is completed; KVO rate is 1ml/h;
9. Power Source: AC100—240V, 50/60Hz; Internal Battery, DC12V car charge;
10. One-key operation makes setup easy and simple;
11. Syringe plunger grabble detector, operable with one hand in a No-Germ Environment;
12. Automatically record the settings of last injection;
Specification of Dual Channel Syringe Pump:
Rate: 50/60ml Syringe: 0.1ml/h-999.9ml/h (0.1ml/h step)
1,000ml/h-1,500ml/h (1ml/h step)
30ml Syringe: 0.1ml/h-900.0ml/h (0.1ml/h step)
20ml Syringe: 0.1ml/h-600.0ml/h (0.1ml/h step)
10ml Syringe: 0.1ml/h-300.0ml/h (0.1ml/h step)
Flow Rate Accuracy: Within ±3% (after correct calibration)
Mechanical precision: Within ±2%
Bolus Rate: 50/60ml Syringe: 1,200ml/h
30 ml Syringe: 720ml/h
20 ml Syringe: 480ml/h
10 ml Syringe: 240ml/h
Purge Rate: 50/60ml Syringe: 1,500ml/h
30ml Syringe: 900ml/h
20ml Syringe: 600ml/h
10ml Syringe: 300ml/h
Volume Limit: 0.1ml-9999.9ml (0.1ml step)
Total Injection Volume: 0.1ml-9999.9ml (0.1ml step)
Occlusion of Dual Channel Syringe Pump:
High: 800mmHg ±200mmHg (106.7kPa±26.7kPa)
Medium: 50/600mmHg ±100mmHg (66.7kPa±13.3kPa)
Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa)
Alarms: Injection soon finish, end of injection, occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc.
Power Source: AC 100V ~ 240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup
KVO rate: 1ml/h
Fuse: F1AL/250/60V, 2pcs inside
Power Consumption: 30VA
IP Classification: IPX4
Equipment Classification: Class II, internal power supply, Type CF
Operating Condition: Ambient temperature: +5℃–+40℃
Transport & Storage Condition: Ambient temperature:-30℃–+55℃
Relative humidity: ≤95%
Dimension: 280mm (L) ×210mm (W) ×130mm(H)